Nitinol Flat Wire — Surface Finish, Tolerance, and Af Verification
For guidewire, stent, endoscopic, and orthodontic engineers and buyers — supplied by ZYTC Alloy, China-based Nitinol manufacturer and exporter

Flat Wire Is Not Just “Round Wire Squeezed Flat”
Rolling a rectangular cross-section introduces non-uniform strain across width and thickness. This affects two things buyers care about: dimensional tolerance consistency and final Af temperature after forming and heat treatment.
ZYTC Alloy supplies medical-grade Nitinol flat wire: thickness 0.02–3.0 mm, width 0.04–16.0 mm.
Surface Finish: A Corrosion and Process Decision
Black oxide (BO). Nickel-enriched surface oxide. In potentiodynamic testing, all BO samples showed breakdown — susceptible to localized corrosion. Must be removed before blood contact or implantation. Also interferes with laser cutting.
Electropolished (EP). Thin, uniform oxide; removes contamination. In ASTM F2129 testing, only 2 of 6 EP samples showed breakdown. The defensible choice for blood-contact surfaces — but EP removes material, so incoming thickness must include the polishing allowance.
Chemically etched (CE). In one study, CE surpassed EP in corrosion resistance, attributed to inclusions on EP surfaces acting as initiation sites. “Electropolished” is not automatically better — the right finish depends on the heat’s inclusion content and the application.
Procurement action: Specify finish by function. Blood-contact: EP with defined removal allowance. Laser-cut: finish compatible with cutting, plus post-cut oxide removal requirement.
Tolerance: The Width-Thickness Ratio Sets the Limit
| Width-Thickness Ratio | Thickness Range | Thickness Tolerance |
|---|---|---|
| 2 < γ ≤ 7 | 0.013 – 0.05 mm | ±0.005 mm |
| 2 < γ ≤ 7 | 0.05 – 0.8 mm | ±0.1T mm |
At low γ (near-square), width spread control is difficult; at high γ (thin, wide ribbon), thickness uniformity across the width is the challenge. ±0.1T is a standard commercial tolerance — for stiffness-sensitive springs, specify the target thickness and allowable range explicitly and confirm capability before ordering.

Af Temperature: Report the Finished Wire, Not the Ingot
Transformation temperature depends on chemistry and thermomechanical history. Flat wire rolling introduces different strain than round wire drawing; any shape-setting or heat treatment shifts Af further. DSC-measured values also do not necessarily match functional transformation temperatures under applied stress — the method must be stated.
Procurement action: Require Af by DSC on the finished flat wire, not the ingot. For body-temperature superelasticity (37°C), specify the target Af range and confirm the method. “Conforms to ASTM F2063” is not sufficient — F2063 covers chemistry and mechanicals; the Af that governs device performance must be specified separately.
Supply Specifications
| Item | Parameter |
|---|---|
| Standard | ASTM F2063 |
| Form | Flat wire / ribbon |
| Thickness | 0.02 – 3.0 mm |
| Width | 0.04 – 16.0 mm |
| Surface finishes | Black, oxidation (gold/tan/blue), clean, bright, electropolished |
| Cross-sections | Rectangular, square, triangular, elliptical, custom |
| Certificate | EN 10204 3.1 with measured chemistry, dimensions, and DSC Af |
Applications
Catheter reinforcement ribbons — directional stiffness at lower profile than round wire.
Endoscopic components — superelastic actuation in confined spaces.
Orthodontic appliances — continuous light force with flat geometry for bracket engagement.
Superelastic springs and actuators — compact designs with controlled force-displacement.
Why ZYTC Alloy
-
Finish matched to function: EP, CE, or bright — with material removal allowance specified for tight-tolerance parts.
-
Tolerance confirmed for your γ ratio: specific width-thickness combination evaluated before quoting.
-
Af reported on finished wire: DSC data for the flat wire, not the ingot.
-
Custom cross-sections: square, triangular, elliptical, special profiles to drawing.
-
China-based manufacturer and exporter: direct mill supply, FOB/CIF, full traceability.
-
Lead time: standard dimensions 7–10 days; custom profiles or finishes 3–4 weeks.
Request a Quote
Please provide: thickness and width (or cross-section drawing), surface finish, target Af range, downstream process (laser cutting / shape setting / direct assembly), tolerance requirement, quantity, and whether DSC data or third-party testing is required.
If the part is blood-contact or implant-bound, state the finish and whether post-processing will be performed — we will confirm the polishing allowance before quoting.
📧 daisy@zytcsteel.com